    LIFEPAK 1000 defibrillator

    LIFEPAK 1000

    Physio-Control is launching a voluntary field action for the LIFEPAK 1000 defibrillator. The company has received reports where customers have attempted to use their LIFEPAK 1000 defibrillator and the device has shut down unexpectedly due to an intermittent connection between the battery and device electrical contacts. The company has determined that this intermittent connection is a result of wear and subsequent oxidation formation between the battery and device electrical contacts. When this occurs, the defibrillator may not be able to deliver lifesaving therapy. The company is contacting customers and advising them to perform Required Customer Actions, including removing and reinstalling the battery from their device(s). The documents below provide detailed information on this Field Action.
     

     